Contraindications to Electrical Stimulation

Stop Sign by Arlette on Flickr

"1. DO NOT apply to the thoracic area (or transthoracically) of a patient with arrhythmia, congestive heart failure, recent myocardial infarction, and other heart conditions

2. DO NOT apply anywhere on the body of a patient with a demand-type implanted cardiac pacemaker or defibrillator or deep brain stimulator

3. DO NOT apply through the carotid sinus area (at the bifurcation of the common carotid artery); it may cause a rise in blood pressure, reflex vasodilatation and slow the heart rate.

4. DO NOT apply transcerebrally (thru the head) at a milliamp level because it may cause changes in brainwave patterns. EXCEPTION Microcurrent can be applied transcerebrally.

5. DO NOT apply through cancerous (malignant) tissue.

6. DO NOT apply through areas of broken or irritated skin. The current flows through breaks in

the skin, causing discomfort. Exception: (This is different from using electrical stimulation for wound healing.)

7. DO NOT apply near or touching protruding metal such as surgical surface staples or external pins because they are excellent conductors of electricity.

8. DO NOT use on any patient who reacts very negatively to the experience or to the sensation of stimulation.

9. DO NOT apply to a patient with undiagnosed pain.

10. DO NOT apply to patients who cannot provide adequate feedback concerning the level of

stimulation (infants, individuals with mental disorders)."

Precautions to Electrical Stimulation

Yield by My Own Worst Nightmare on Flickr

"1. USE CAUTION in applying at high amplitude directly over areas where bone is superficial. Periosteal pain can result.

2. USE CAUTION when applying in areas of excessive adipose tissue since the high levels of stimulation necessary to activate underlying structures may cause pain or autonomic reactions

3. USE CAUTION in applying within 3 feet of a transmitting cellular phone or two-way radio. This may cause electrotherapy equipment malfunction.

4 USE CAUTION in applying near the uterus during pregnancy and delivery. (The possible effects on a fetus are not known.)

5 USE CAUTION in applying within 10 feet of Group 2 ISM equipment that generates high frequency or high energy electromagnetic radiation. Such equipment includes welding or cutting equipment, diathermy units & surgical electrocautery units. This may cause electrotherapy equipment malfunction.

6. USE CAUTION when applying in the region of the urinary bladder because the current may interfere with normal function

7. USE CAUTION when applying over scar tissue because the scar will have an increased electrical resistance. The current will preferentially travel around the scar causing increased current density at the edges of the scar with possible burning.

8. USE CAUTION in applying to a patient with history of metastatic disease.

9. USE CAUTION in a patient with poor sensation."